English
 
Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  Entwicklung einer Kalibriereinrichtung und eines Sensors für magnetische Feldmessungen in einem Frequenzbereich von 100Hz bis 1MHz

Pulz, E. (2018): Entwicklung einer Kalibriereinrichtung und eines Sensors für magnetische Feldmessungen in einem Frequenzbereich von 100Hz bis 1MHz, (Scientific Technical Report STR ; 18/01), Potsdam : Deutsches GeoForschungsZentrum GFZ.
https://doi.org/10.2312/GFZ.b103-18012

Item is

Files

show Files
hide Files
:
STR_1801_Pulz.pdf (Publisher version), 5MB
Name:
STR_1801_Pulz.pdf
Description:
Fulltext
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
CC-BY-SA 4.0
:
STR_1801_Pulz_a.pdf (Publisher version), 5MB
 
File Permalink:
-
Name:
STR_1801_Pulz_a.pdf
Description:
Archive version
Visibility:
Private
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
CC-BY-SA 4.0

Locators

show

Creators

show
hide
 Creators:
Pulz, Eberhard1, 2, Author              
Affiliations:
12.3 Earth's Magnetic Field, 2.0 Geophysics, Departments, GFZ Publication Database, Deutsches GeoForschungsZentrum, ou_146030              
2Scientific Technical Report STR, Deutsches GeoForschungsZentrum, ou_9026              

Content

show
hide
Free keywords: -
 Abstract: A new calibration facility was developed, tested, and installed at the premises of the Geomagnetic observatory in Niemegk to test induction coil magnetometers (search coils) in the frequency range 100Hz to 1 MHz. The new calibration facility is based on a Braunbeck coil, which extends existing installations based on a three-dimensional Helmholtz coil and a cylindrical coil which cover only the frequency range from a few mHz to 100 kHz. The new calibration coil is primarily needed to test magnetic field sensors used by the Geophysical Instrument Pool Potsdam for the so-called Radio Magnetotelluric (RMT) method. The RMT method uses electromagnetic fields transmitted by radio stations to infer the electrical conductivity distribution of the shallow subsurface (~1 m to 50 m). We present theoretical calculations to determine the parameters of the new calibration coil, which could be confirmed with measurements using an OMICRON BODE 100 spectrum analyser. The new coil system was also tested by calibrating two existing magnetic field sensors (METRONIX) MFS07 and SHFT-02e) with known response functions. We infer an overall accuracy of the new system of <0.05nT in amplitude and <0.5° in phase for frequencies <700kHz and ≤0.05nT and ≤3° degrees for frequencies > 700kHz. We also developed, tested and calibrated a new magnetic field sensor, using a set of three orthogonal toroidal coils (based on the antenna AN 200). The initial results look promising. The overall noise figures surpass those of the Metronix SHFT-02e sensor. The amplitude characteristic is slightly non-linear though, varying between 5 and 75 mV/nT over the frequency range 10 kHz to 1 MHz.

Details

show
hide
Language(s): deu - German
 Dates: 2018
 Publication Status: Finally published
 Pages: -
 Publishing info: Potsdam : Deutsches GeoForschungsZentrum GFZ
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.2312/GFZ.b103-18012
URN: urn:nbn:de:kobv:b103-18012
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Scientific Technical Report STR
Source Genre: Series
 Creator(s):
Affiliations:
Publ. Info: Potsdam : Deutsches GeoForschungsZentrum GFZ
Pages: - Volume / Issue: 18/01 Sequence Number: - Start / End Page: - Identifier: ISSN: 2190-7110